SPECTROPHOTOMETER
The PASTEL Uvi-Line is an innovative instrument that allows to perform analysis also without reactive.
In mode “with reagents” it works as a standard UV-Vis spectrophotometer as good as UviLine 9400; in mode “without reagents” it can simultaneously measure COD, BOD, SST, TOC, and the concentrations of nitrates and anionic surfactants: this saves a lot of time in respect to the standard analytical methods. This process is based on the spectral deconvolution that allows it to express quantitative and qualitative results about the analysed sample. This option is operational only in UV field and there are concentration limits beyond which the measurement can not be carried out. The time required for the results is just 2 minutes.
Upper and lower limits in mode “without reagents”:
TEST | RANGE |
Total Suspended Solids (SST) | 2,5 – 100 mg/L |
Chemical oxygen demand (COD) | 1 – 500 mg/L |
Biological oxygen demand (BOD) | 0,25 – 350 mg/L |
Total Organic Carbon (COT) | 0,5 – 300 mg/L |
Nitrates (NO3-N) | 0,025 – 25 mg/L |
Anionic Surfactants | 0,05 – 60 mg/L |
This instruments is available with different holder: for 16mm round tubes, for 10mm square cuvettes or with an universal holder for 16mm round tubes, 10mm square cuvettes and 5 mm, 20mm, 50mm and 100mm rectangular ones.
Ready to use in less then 3 minutes from switching on, it has many options for customization beside a wide range of languages to choose from.
TECHNICAL SPECIFICATIONS | |
Wavelength | 190 – 1100 nm |
Light source | Xeno |
Band width | 4 nm |
Wavelength accuracy | ± 1 |
Wavelength repetability | ± 0,2 nm |
Absorbance range | -3 / 3,3 A |
Absorbance resolution | 0,001 |
Absorbance accuracy | 0,3% a 0,003 Abs < 0,600 Abs |
Stray Light | < 0,1% a 220, 340 e 400 nm |
Interface | 1 USB-A / 1 USB-B / 1 RS232C |
Operating temperature | 10 – 35°C |
Power supply | 110 – 220 V |
Dimensions (LxWxh) | 404 x 314 x 197 mm |
Weight | 4 Kg |
Cell holders | 16 mm Tube and 5 to 100 mm Cuvette |